About Us | Help Videos | Contact Us | Subscriptions
 

Agronomy Journal Abstract -

An Inexpensive Silver-Silver Chloride Reference Electrode

 

This article in AJ

  1. Vol. 88 No. 2, p. 246-249
     
    Received: Jan 26, 1995


    * Corresponding author(s): lshuman@gaes.griflin.peachnet.edu
 View
 Download
 Alerts
 Permissions
Request Permissions
 Share

doi:10.2134/agronj1996.00021962008800020021x
  1. Larry M. Shuman ,
  2. Michelle M. Wander and
  3. Ray B. Pitts
  1. Dep. of Crop and Soil Sciences, Univ. of Georgia, Georgia Exp., Griffin, GA 30223

Abstract

Abstract

Simple, inexpensive, and rugged reference electrodes are essential for field applications. The objective of this study was to develop an inexpensive, field-worthy reference electrode. The electrode developed consists of a silver wire coated with AgCl held in a small glass tube filled with 4 M KCl gelled with agar. The glass tube assembly is fitted inside a larger polyvinyl chloride (PVC) tube filled with gelled (NH4)2SO4 lution and capped with a fritted soil-contacting membrane. The millivolt output compared well (R2 = 0.999) with predicted values and values from commercial reference electrodes when paired with pH and nitratesensing electrodes. Comparisons of commercial and fabricated electrodes in two soils showed that the outputs were statistically the same, and that they were operational within a moisture range from saturation to 0.7 to 0/1 kg kg−1, depending on the soil texture. The fabricated electrodes are low cost, rugged, and stable for months when kept wet.

  Please view the pdf by using the Full Text (PDF) link under 'View' to the left.